Iimitha zokuhamba kwe-Ultrasonic

Amava okuvelisa angaphezu kweminyaka engama-20

xa umlinganiselo wexesha oboniswa kwi-M91 udlula uluhlu lwe-100±3%, (Eli lixabiso lesalathiso nje), umsebenzisi kufuneka ajonge:

1) Ukuba iiparameters zombhobho zingeniswe ngokuchanekileyo.
2) Ukuba isithuba sokufakelwa ngokwenene sihambelana ngqo nexabiso le-M25.
3) Ukuba ii-transducers zifakwe ngokufanelekileyo kwicala elifanelekileyo.
4) Ukuba ubude bombhobho othe tye buyanele.
5) Ukuba umsebenzi wokulungiselela wenziwe njengoko kuchaziwe apha ngasentla.
6) Ukuba umbhobho ugqwethekile okanye kukho ukungcola okuninzi ngaphakathi.
Qaphela: Ukuqinisekisa ukuba i-1-6 ayiyongxaki, uze emva koko ulungise umgama we-transducers ukuya kwiqiniseka ukuba iTOM/TOS=97-103% kwimenyu91. Okukhona isondele kwi-100%, kokukhona ingcono.
I-Whatsapp:+86 189 1792 0598

Ixesha lokuthumela: Matshi-07-2022

Thumela umyalezo wakho kuthi: